Sound Power long manual

For more information about Order tracking take a look at the Sound power manual.

Sound Power short manual

The DewesoftX Sound Power plugin calculates sound power from microphone sound-pressure measurements. It accounts for the measurement surface, background noise, room influence, and meteorological conditions.

Results include overall sound power, octave and narrowband spectra, time histories, averages across runs, and apparent directivity. The separate Order Analysis Sound Power module calculates sound power by rotational order.

1. Installation and microphone calibration

Install the Sound Power plugin and activate valid DewesoftX and Sound Power licenses. A 30-day evaluation license is available. Enable the plugin under Settings → Settings → Extensions. Refresh the extension list if the plugin does not appear.

Connect the microphones and set their physical quantity to Sound Pressure in Channel Setup. Calibrate each channel by entering the sensitivity from its calibration certificate, or fit an acoustic calibrator and select Calibrate to determine sensitivity automatically.

2. Standard, geometry, and microphones

Select the measurement standard, then enter the dimensions and choose the microphone arrangement.

Standard selection Available configuration
ISO 3741 Reverberation-room measurement with user-defined microphone positions
ISO 3743 Sound-pressure-based measurement with user-defined positions
ISO 3747 Comparison method in reverberant environments
ISO 3744 Parallelepiped, cylindrical, or hemispherical arrangements
ISO 3745 Hemispherical or spherical arrangements
ISO 6395 / ISO 6396 Hemispherical arrangements for earth-moving machinery applications

For ISO 3744, select the device size and installation condition: on a floor, beside one wall, or in a corner. Enter the reference-box dimensions and measurement distances. For spherical and hemispherical surfaces, enter the measurement radius. The available geometry fields depend on the arrangement.

Choose the applicable microphone-position table and assign an input channel to each position. The position list shows microphone coordinates and group assignments. The plugin warns if the microphone count differs from the selected standard.

The standard also sets requirements for the test environment, instrumentation, operating conditions, and measurement uncertainty.

Microphone grouping

If you have fewer microphones than required positions, divide the positions into groups. Measure one group, move the microphones to the next positions, and continue. The plugin combines the group measurements into the final result.

Sound Power configuration in DewesoftX, showing microphone geometry and position visualization

Measurement geometry and microphone channel assignments.

3. Spectral analysis and measurement timing

Setting Function
CPB resolution Calculate 1/1-octave or 1/3-octave sound power spectra.
Narrowband FFT Calculate FFT results alongside CPB spectra to examine tones and individual frequency components.
Bandwidth Set the lower and upper analysis limits. Available limits depend on the acquisition rate.
Acoustic weighting Select A, C, or unweighted Z results.
Averaging duration Set the acquisition interval used for the averaged result.
Measurement delay Delay averaging after a measurement starts.

Choose the averaging duration according to the selected standard and the source’s operating cycle.

Start and stop triggers

Enable Use trigger channels to control acquisition from external signals such as light barriers. Select the start and stop channels, edge directions, trigger levels, and retrigger levels.

You can use one channel with different start and stop conditions, or separate channels. Acquisition starts at the first qualifying start edge and ends at the first qualifying stop edge.

Parallel CPB and narrowband FFT sound power analysis in DewesoftX

CPB and narrowband FFT results calculated in parallel.

4. Correction factors

The plugin calculates sound power from the sound-pressure result, measurement surface, and applicable corrections:

Lw = Lp + 10 log₁₀(S / S₀) − K1 − K2 + C1 + C2

Here, S is the measurement surface and S₀ = 1 m². Select the correction methods required by the measurement procedure.

K1: background noise

Choose one of four methods:

  • Calculate from measured data: acquire background noise during the test. You can configure multiple background runs and add a reference curve for comparison.
  • Enter values: enter previously determined K1 corrections.
  • Enter background noise: enter background-noise levels and let the plugin calculate K1.
  • From previous measurement: reuse the last acquired background-noise measurement.

Set the analysis bandwidth before entering correction data. Changing bandwidth reconfigures the tables and can clear entered values.

K2: room or location influence

Method Required information
Mean absorption grade Room dimensions and mean absorption grade
Reverberation time Room dimensions and measured reverberation time, for example from the RT60 module
Enter values Previously determined K2 corrections
Calculate from RSS Certified reference sound source levels and a measurement of that source in the test environment

For the RSS method, enter the source’s specified Lw (RSS) values in the editor. The plugin calculates K2 using the measured Lw* values. Open the K2/RSS editor to view or edit these tables.

C1 and C2: meteorological corrections

Enable meteorological correction and enter the temperature and either pressure or altitude. Entering pressure calculates altitude, and vice versa. These corrections apply to the conditions described in the full manual: temperatures below 23 °C or test-site altitudes above 500 m.

Sound Power correction settings for background noise, room influence, and meteorological conditions

Background-noise, room, and meteorological correction settings.

5. Measurement and live results

Open Measure to use the automatically generated SoundPower display. It contains acquisition controls, microphone-group selection, progress messages, status, and warnings.

The input meters and spectra show microphone sound-pressure data. The result meters and plots show overall sound power, combined sound pressure, spectra, and correction factors.

For a typical grouped measurement:

  1. Start storing with the test source switched off.
  2. Acquire background noise for the active microphone group.
  3. Switch on the source and acquire the source measurement.
  4. Move the microphones to the next group and repeat the required background and source acquisitions.
  5. Complete all groups and runs, then check the results and warnings.

Use the previous and next group controls to move through the sequence. You can repeat a group measurement before the full acquisition sequence finishes. In the Runs editor, you can place background and room-correction measurements before or after source runs.

6. Runs, time histories, and directivity

Multiple runs and operating modes

Enable Number of runs to combine repeated measurements or different operating modes. The averaged overall, CPB, and enabled FFT results update after each completed run.

Use the Runs editor to rename runs, change their order, and set their relative weights. All runs have equal weight by default.

Enable Show history channels to retain the individual-run sound power results. These can be displayed in a table with their run identifiers. The Results Matrix contains individual microphone results and averages across runs, and can be exported for further analysis.

Sound power versus time

Enable Use Lw vs. time output and choose an averaging interval to generate:

  • Overall sound power versus time.
  • A CPB sound power waterfall versus time.
  • An FFT sound power waterfall versus time, when narrowband processing is enabled.

Use these outputs to examine changes during operating cycles and transient events.

Apparent directivity

Enable the Apparent directivity index to compare sound radiation at each microphone position with the mean across the array.

The plugin calculates overall and octave-band directivity for each microphone. It also outputs the maximum apparent directivity and the identifier of the microphone at that maximum.

7. Offline calculation

Store raw microphone signals to calculate sound power after acquisition or revise an existing calculation. You can change supported analysis settings, including frequency weighting and bandwidth, offline. Microphone grouping cannot be changed.

Calculation mode Data selection
Lw between locked cursors Calculate sound power over the interval between L1 and L2.
K1 and/or Lw from locked cursors L1 marks the background-noise start; L2 marks the source-measurement start. The configured averaging duration sets the length of each interval.
K1, K2, and Lw from events in selected time region Recalculate using acquisition events within the selected region. The Sound Power plugin must have been active during recording.

Open the recording, configure Sound Power under Offline math, select the calculation mode, and return to Review. For cursor modes, position and lock both cursors before selecting Recalculate.

When recording raw data for later processing, include enough background-noise and source-running data to cover the intended averaging duration.

8. Order Analysis Sound Power

Order Analysis Sound Power (OASP) calculates sound power by rotational order across a speed range. It can help identify noise from rotating components, including blade passing, gear meshing, and engine firing.

OASP requires both the Sound Power and Order Analysis licenses.

  1. Configure Order Analysis with microphone sound-pressure inputs and a rotational angle or speed reference.
  2. Add Order Analysis Sound Power from the Acoustics application modules.
  3. Select the Order Analysis instance that will supply the data.
  4. Enter the radiating measurement-surface area, temperature, and barometric pressure. Exclude reflecting surfaces from the entered area.
  5. Select linear or A-weighting. Check the microphone contributions and extracted orders in the Positions table.

If you change the source Order Analysis configuration, use Add to update the OASP inputs.

Results include a sound power order waterfall versus speed and individual extracted-order curves. Graph markers provide sound power values at selected operating points. The reference axis follows the configuration of the source Order Analysis instance.

Order Analysis Sound Power setup and results, including order waterfalls and extracted-order curves

OASP settings, an order waterfall, and extracted-order curves.

9. Export, print, and reports

Export results to formats such as Excel, CSV, MATLAB, DIAdem, and RPCIII. To copy data from an individual display instrument, select the instrument and copy its data to the clipboard.

Use Print to print the current display or create a PDF through a PDF printer.

Excel report template

The supplied Sound Power report combines measurement information, overall levels, and CPB results in a customizable Excel workbook.

To set up reporting:

  1. Install Microsoft Excel.
  2. Import SoundPowerDataHeader.xml under Settings → Settings → Data Header and choose when to request measurement details.
  3. Copy SoundPowerReport.xps, SoundPowerReport1.xlt, and SoundPowerReport1.xltx from the SoundPower add-on folder to the DewesoftX Scripts folder.
  4. Restart DewesoftX, open the measurement, and select the Sound Power Excel export template.
  5. Export Single value channels to populate the report without exporting unnecessary full-speed signal data.

Select Edit template to change the logo, layout, or linked result cells. Before saving the reusable template, clear generated content from sheets other than Results, as described in the full manual.

Example Sound Power report generated from DewesoftX results

An Excel report with measurement details, spectra, and final results.

10. Automated workflows

Use the DewesoftX Sequencer to automate supported Sound Power operations. Add an Action block, select SendPluginCommand, and enter the Sound Power plugin identifier:

{09FC95F3-18B1-41AE-A44E-DD4D0719614A}

Command IDs Function
2000 to 2002 Configure offline processing, cursor-based calculation, and unweighted or A-weighted analysis
2010 to 2015 Import K1, background-noise, K2, RSS reference/measured data, and RSS bounds
2020 to 2025 Export the corresponding correction and reference data

These commands let you reuse correction data, back up measurements, and repeat post-processing procedures. Each command has specific inputs and operating-mode requirements.

Download the complete Sound Power manual for detailed procedures, examples, command parameters, and safety information.